Photographing across the depth of a thin section
One way to assess biological tissue quality is to take photographs
across the width or length of a thin section. The photographs are taken of
every field or every other field, for example, in a straight, line direction.
The ability to rotate the grid, via a rotating holder, orients the grid for
this type of photography. See diagrams below depicting Epon thin sections with
either 400 or 100 um tissue slices. Each thin section is oriented for the
straight, line photography.

The photography should begin at the top or to one side of
the thin section. By moving the X or Y stage controls on the electron
microscope, every field in a straight line can be photographed. On the JEOL 1200 EX and 2010 electron
microscopes, it is the field within the outer brackets, marked on the
phosphorescent screen, which is included in each photo/negative. See diagram
below. Note that the photographic field is rectangular. In addition, the label
projected onto each negative is usually located on the right side of this
field. The photographic field of the same area will naturally encompass more or
less when decreasing or increasing magnification, respectively. For example, at
approximately 6000X, the photographic field will include an area 12.5 um x 16.2
um. At approximately 10,000x, the
photographic field will include an area 7 um x 9 um. Consequently, this
rectangular field is used as a ruler to measure distances.

The scale bar adjacent to the brackets on the screen is
equal to 5 cm. At approximately 10,000X, this scale bar is equivalent to 5 um.
On the JEOL 2010, moving the X, Y stage controls shows a corresponding readout
on the CRT of the distance in um traveled. These are all the tools that are
available to take photographs in a straight-line fashion.
In our
laboratory, we photograph a thin section of the rat hippocampus, across the
length or width, from one surface to the other. In some studies, it is of
interest to study the tissue near the vicinity of electrode placements. This
latter interest will serve as an example of such photography.
Procedure:
- At LOW
MAG mode, e.g. 60X, orient the grid by rotation so that the air versus net
will be in the north/south orientation, respectively. Study the appearance
of the tissue/section so that different areas will still be recognizable
even though the magnification will be increased, e.g., to 6000X, for
photography. If the electrode hole is obvious, position the central black
dot on screen near it.
- Put
the objective aperture back in the beam path and increase the
magnification to 6000X. The area near the electrode hole should be in
view. Rotate the grid as needed to re-orient the tissue.
- Align
the length or width of the rectangular photo field, depending on
preference, to the top of the tissue. Rotate the grid as needed. See
diagram:

- To
determine if the sequence of photography can be maintained in a relatively
straight line, from “air” surface to “net” surface, turn either the X or Y
stage control, whatever is applicable. Continue turning, until the entire
field, from top to bottom in this case, is traversed.
- Switch
to LOW MAG mode again, remove objective aperture, and observe where the
central black dot (on the phosphorescent screen) ended up. If it is
located directly across from the starting point, then the grid is oriented
properly. If it is not, then try rotating the grid in clockwise or
counterclockwise until the orientation is acceptable.
- Put
the objective aperture back in. Go back to the air surface at 6000X, and
prepare to photograph across the depth of the tissue.
- The
first step is to align the rectangular field (either with its length or
width) to the “air” surface and rotate the image as needed. Include a bit
of Epon interface to indicate the “air” surface in the first photograph,
if so desired.
- To
take an electron micrograph, first, focus the image using the wobbler as
needed.
- Once
in focus, one can under-focus 2 clicks on the fine focus knob on the JEOL
1200 EX and on the JEOL 2010, one can set the auto-under-focus to 2. These
values are arbitrary. You should first take a series of photos at all the
under-focus settings and determine which works the best for your research.
Leave the small screen in place during the photo shoot.
- Set
the brightness or illumination to 0.5 seconds exposure. This short
exposure seems to work best on sections attached to coated slot grids
because there is less time for image drift to occur. If there is image
drift present, then wait a few minutes before taking a photo. Hopefully,
with time, the drift becomes minimal.
- Check
that the photo setting is on AUTO and not MANUAL. (The MANUAL setting is
good for taking photos under LOW MAG conditions where there is a range of
exposure times to choose from).
- Make
sure that the all white incident lights are off in the room. Lock the room
door to avoid someone walking in during a photo shoot.
- Press
the photo button once or twice, as needed, to shoot the picture. Avoid
touching the scope during the 0.5 seconds of photography to avoid
vibrations. Once the photo is shot, prepare to move on to the next field.
- Turn
the X or Y stage control (whatever is appropriate) to move to the adjacent
field. Alternatively, one may choose to shoot every other field at
pre-calculated depths. Remember that the rectangular photo field is used
as a counting device of um traversed. See diagram:

- Once
all photographs have been taken, it is useful to return to LOW MAG mode
and observe the path of photography traversed, indicated by the beam
“burn”. If all the photos followed a relatively straight path, the area
burnt in by the beam is obvious. Take a photograph of this image under LOW
MAG.
- To
take a photo under LOW MAG, focus on the image as usual, using the wobbler
assistance.
- Set
the photo to MANUAL.
- For
60X to 100X, set the exposure toggle to either .35 sec or .5 sec exposure
(the second setting will result in a darker negative). For 100X to 200X,
set the exposure to .5 sec or .7 sec.
- Leave
the small screen in place.
- Press
photo button once or twice to photograph the image.
- Now
you have a record of the depth profile photographed. This low
magnification image of the electron beam “burn” markings is also useful
when isolated photos are taken at a distance from the tissue surface. For
example, if only several photos are taken sequentially in a straight path
across a tissue, then the positive print or negative of the low
magnification of the beam burn markings is a useful tool. One can measure
distances in um on the print or negative itself! See diagram:
