Loading and Unloading the
Gatan Rotational Holder into the JEOL 2010 Electron Microscope
4/11/01
Materials:
JEOL 2010
Electron Microscope, Gatan Single Tilt Rotational Holder Model 650, motor drive
control unit, mouse control, connecting wires with plug.
Procedure: Inserting the Gatan Rotational Holder
- Follow
the diagrammed instructions for rotational holder insertion into the
specimen chamber. Go to page 3 on the TV screen to observe the vacuum
sequence. See diagram below:

Note that the rotational holder (RH) must be manually supported
until the vacuum pull is strong enough to self-support in step #5. It usually takes several minutes for the
rotational holder with a glued grid/hex ring to pump down sufficiently until
the green light comes on (step #4). Just be patient!
- Prepare
to plug in the connecting cord into the handle of the RH. Note the contour
of the plug jacket and the socket. See diagram below.

Note also that the plug has an
outer sleeve that must be pulled back. Note the corresponding contour of the
receptacle in the handle as well as the configuration of the prongs. Carefully
connect the plug into the receptacle.
- If the
insertion of the RH ever causes the vacuum to shut down, then suspect that
the two black O-rings on the holder may need some greasing. In this case,
remove the holder from the electron microscope. Then remove the two
O-rings. Note that one is smaller than the other. With a Kim-wipe, clean
each O-ring with 70% ethanol. Allow them to air dry. Then, smear a small
amount of white grease on each ring. Wipe away excess with a Ross lens
tissue. Next, re-attach the correct size O-ring to their positions on the
RH. Finally, try re-inserting the RH into the electron microscope again.
- Once
the specimen is in the electron microscope, set the MAG mode to LOW MAG,
without any objective apertures in the way.
- Turn
on the filament current to view the grid at 50X or 60X for several
minutes. At the same time, exposure of the Pioloform coated grid to this
low level electron beam current will serve to “condition” the entire grid.
This “conditioning” should result in a more stabile Pioloform structure
and consequently, less image drift.
While waiting, one can count the number of sections in the ribbon,
also noting the location, orientation and disposition of the sections. One
can also photograph the ribbon at LOW MAG.
Procedure: Removing the Gatan Rotational Holder
- Follow
the diagrammed instructions for removing the holder from the column.

- Carefully
unplug the connecting cord from the holder.
- Return
rotational holder to stand.