Thick Sectioning the Epon Block
When using the ultramicrotome for sectioning, remember to
turn on the compressed nitrogen to the air table.
- Following trapezoid
trim, place the Epon block, which is still in the chuck, into the specimen
arm of ultramicrotome and secure tightly.
- Set the block/specimen
arm angle to 0 degrees.
- Place a glass knife
into the knife holder and secure tightly. This glass knife will be used to
“rough face” the block. Rough facing means removing most, if not all the
Epon lying over the tissue in an Epon block in um increments on the
ultramicrotome. A glass knife is used for this job because it is
inexpensive and replaceable.
- Set the knife clearance
angle between 4 to 6 degrees.
- Set the knife stage
angle to 0 degrees.
- Set the cutting speed
to no more than 1 mm/sec on the Ultracut T. For the Ultracut E, a faster
speed can be used (up to 3 mm/sec).
- Set the section
thickness to no more than 1 micron (um) on the Ultracut T. The Ultracut E
can be more accommodating and can handle up to 3-um increments.
- Advance the left edge
of the glass knife to within several millimeters (mm) of the block face.
The left edge of a glass knife is the straighter and better edge of the
knife to use in thick- and thin sectioning. Look through the
stereomicroscope/binoculars to judge this distance approximately. Then
lock the stage securely.
- Set the length of the
cutting window. To do this on the Ultracut T, note the “window” area of
the control panel. There is an upper and lower button.
- Hand rotate the hand
wheel clockwise to position the block several mm above the level of the
knife-edge. Then depress the upper
button once.
- Position the block
several mm below the knife-edge by rotating the hand wheel clockwise. Then depress the lower button
once. NEVER ROTATE THIS HAND WHEEL COUNTERCLOCKWISE, UNLESS IT IS
FOR A FEW MILLIMETERS ONLY!!! THIS HAND WHEEL CAN ACCOMMODATE A FEW
MILLIMETERS COUNTERCLOCKWISE TURN.
- To check whether this
is an acceptable length of the cutting stroke, turn on the motor to the
ultramicrotome and observe the cutting motion. The cutting stroke should
commence when the block is several mm above the knife-edge. The end of the
cutting stroke is denoted when the entire block face is several mm below
the knife-edge. Repeat steps #9 through #12 for the Ultracut T if
re-adjustments are needed.
- To set the length of
the cutting window on the Ultracut E, firstly, lock the hand wheel in
place. Once the hand wheel is in
the locked mode, turn the hand wheel a bit clockwise or counterclockwise
to bring the block below the level of the knife. Then set the desired
length of the cutting window in mm by adjusting the lever on the hand
wheel. Unlock the hand wheel. Turn
on the microtome and observe the cutting motion. If the length of the
cutting window is longer than needed or too short, then turn off the
microtome. Change the cutting
length by locking the hand wheel again, then adjusting the lever on the
hand wheel to a different value in mm.
Turn the microtome on again and assess the new cutting length. When
finished, turn off the microtome.
TWO IMPORTANT RULES IN MICROTOMY:
1) ALWAYS RETRACT THE KNIFE
AWAY FROM THE BLOCK FACE WHEN MAKING ANY ADJUSTMENTS TO AVOID ACCIDENTAL
COLLISIONS.
2) PROPERLY TIGHTEN AND SECURE
ALL COMPONENTS OF BLOCK HOLDER, KNIFE HOLDER AND STAGE HOLDER TO ENSURE SAFE
AND PROPER USAGE OF THE ULTRAMICROTOME.
- The next step is to
align the block face with the knife-edge (left edge of the glass knife).
For now, the block sits in a 0-degree angle, as well as the knife stage.
We will assume, for now, that the tissue is oriented perfectly
perpendicular to the knife. Turn off the fluorescent overhead light in the
microtome and use the bottom light source only. Advance the knife-edge
close to the block face. Increase the magnification up to 4X. The
knife-edge, when brought into close proximity with the block face, will
cast a reflection onto the block face. The closer the proximity, the
narrower the white shadow becomes. This occurrence is what makes an
alignment possible. Regardless of
whether the block face is uneven, due to razor trimming or smooth and
even, there will be some degree of a white shadow seen. We use this shadow
to judge how distant or close the areas of the block face are to the
knife-edge. If block and knife-edge are in perfect alignment in terms of
the knife stage angle, then the white shadow appears as a thin, even,
horizontal white line of light (horizontal alignment). See diagram of
horizontal alignment in step #21. If the block face is in perfect vertical
alignment, then as the block is put through the cutting motion (manually
rotate the hand wheel clockwise), the white shadow reflection continues to
appear as a thin white line. If
the vertical alignment was off, this means the top or bottom of the block
is to close or too far from the knife-edge. Then the block angle must be
adjusted to a setting other than 0 degrees.
- As stated, we will
proceed with the block/stage being in the 0,0 degree setting. Therefore,
we will “face” or trim the block face as it is. To do this, bring the
knife-edge as close as possible to the block face without colliding into
the block face at any point.
Rotate the hand wheel slowly to see where any vertical misalignment
lies. If the top or bottom of the block is protruding, then this area will
inevitably be cut off first.
- Then, turn on the motor
and allow the microtome to remove the overlying Epon in 1 um (or more)
increments. Continue to face or trim the block until entire Epon sections
are removed. Or stop, once some tissue is seen being cut. Then, retract the knife away from the
block face. Use a small vacuum to remove Epon trimmings on and around the
microtome.
- The tissue in the block
should now be ready for thick sectioning. It may be necessary to re-trim
the block to a smaller trapezoid if the block face has gotten too
large. Remove the block from
specimen arm and with a new razor blade, trim any surrounding Epon, close
to the border of the tissue. Or
leave a larger margin of Epon around the tissue, depending upon the nature
of the experiment. The resulting trapezoid should be no wider that 3 mm at
the base and no taller than 3 mm in height. The smaller the trapezoid, the
easier it is to section. Return
the block to the specimen arm.
- For good thick
sections, switch to a new glass knife with a boat attachment or a diamond
Histo knife. New students are to
continue to use glass knives for all sectioning.
- Place a new glass or a
diamond Histo Knife in the holder and secure it tightly. Set the clearance
angle between 4 to 6 degrees, generally 6 degrees is most often used. Add enough twice-distilled water to
fill the knife boat to make it level with the knife-edge. It is advised to
pre-fill the diamond knives with water about ½ hour before use in order to
wet down the cutting edge for the sake of convenience. If the diamond
knife-edge repels water, then add a few saliva drops to the dry knife-edge
for several minutes. This seems to
render the diamond knife-edge hydrophilic. Then rinse the knife well under
running 2X distilled water. To dry, carefully wick away excess water
droplets with filter paper. Avoid
touching the cutting edge with anything other than an eyelash hair or fine
brush hair. Secure the diamond knife in holder.
- When using a glass
knife to thick or thin section, some kind of boat must be attached to the
knife. See boat making for glass knives in the protocol entitled, “Glass
Knife Making”.
- Align the new knife to
the block face. Look through the binoculars while carefully advancing the
knife to vicinity of the block face. It is useful to increase the
magnification to up to 4X and use only the bottom light source. Again,
note that the knife casts a shadow onto the block face. We will now check
the horizontal alignment. The block face surface should now be very even
and more reflective. As the knife is advanced even closer, the shadow
should appear as a thin white horizontal line across the block face. See
left diagram below for a perfect horizontal alignment. The diagram on the
right indicates, on the other hand, that knife-edge is too close to the
left side of the block surface. Remember to retract knife from block
before making any adjustments. Then, to adjust, rotate the knife stage a
few degrees to the right. Then, check alignment again. Continue to adjust
knife stage as needed.

- Check the vertical
alignment. Re-advance the knife to
the block face to within a mm or more, while looking through
binoculars. Then slowly rotate the
hand wheel clockwise to pass the entire height of the block near the
knife. Because the block has already been “faced”, the vertical alignment
should be fine.
- Now, prepare to thick
section the block. Re-check the length of cutting window. Set section
thickness setting to 1 um or less. Set the cutting speed to 1 mm/sec.
Advance the knife close enough so that the white line shadow becomes a
green line. Any closer and the line becomes red. This is close
enough. Turn on the microtome and
wait for 1 um sections (or less) to begin cutting.
- The first sections may
contain partial tissue and partial Epon. Or the surface tissue may be considered
undesirable. Or every section containing tissue is critical. Stop the
microtome when enough sections are cut. Remove any unwanted sections from
the boat by picking up with a clean, fine artist’s brush. Position the
brush under a section(s) and lift out of the water. Save the good sections
by transferring them on the brush, one at a time, to a glass microscope
slide. Several sections can be placed on a large drop of water on the
slide. See diagram of left slide. Or place each section on individual tiny
drops of water. As a suggestion,
place the deepest cut section apart from the rest, perhaps at the extreme
right side on the glass slide. See diagrams.

27.
Once
all the desirable sections have been placed on glass slides, transfer the
slides to a hot plate for drying. Set
the hot plate temperature hot enough so the sections dry completely in 15 to 20
minutes. If they show too many wrinkles, then they are drying too fast due to
too high a temperature setting. If the hot plate is not hot enough, then the
sections will not adhere to the slide and will inevitably be rinsed off during
staining. To remove unwanted floating sections or debris on the boat water,
wipe the surface water with a small piece of Ross Lens Tissue.
28.
The
dried sections should be stained with 1% Toluidine Blue. To stain the sections, apply filtered
Toluidine Blue to the sections following their dry down on the hot plate. The
stain is 1 part 70% ethanol to 1 part 1% Toluidine Blue. Stain for 30 sec to 2
min, depending on how dark the sections should be. An iridescent rim forms and
the stain begins to “smoke” as it begins to evaporate. Remove slide from hot plate and allow it to
cool for one minute on the countertop.
29.
Rinse
the sections over a waste container, first with a gentle stream of 2X-distilled
water, followed by a stream of 70% ethanol. Lastly, rinse the sections with the
water again. Shake off the excess water. Wipe the back of the slide with a Kim
wipe, then place the slide back on the hot plate to dry.
30.
Remove
the dried slide from the hot plate and place it on a paper towel on the counter
to cool to room temperature. Write in pencil the name of block/tissue, slide #,
today’s date, etc., on the label end of slide.
Prepare to coverslip it.
31.
To
coverslip, add 3 to 4 drops of Permount mounting media across the length of the
portion of the slide containing the sections. Place one drop on the label end.
32.
Slowly
lower a glass cover slip (50 mm x 24 mm) at an angle over the section portion
of the slide. Avoid trapping any air bubbles in the process. Place a smaller
coverslip (22 mm x 22 mm) over the label end.
See diagram:

33.
Place
the coverslipped slides in a dust-free area (in a box, e.g.), to dry as is for
several days. Then store permanently in a slide box.
34.
Alternatively,
one may wish to examine the slides under a light microscope before
coverslipping to examine the stained sections.
Studying the section, you may wish to cut deeper into the block for more
sections. Or you may not like the present angle of the block if the sections
look obliquely cut. Then you must change the block or knife angle or both. If you are keeping track of the length of
the tissue then you may wish to remove the block from the microtome and re-measure
it. (See Special Procedures at end of this protocol). Whenever a block is
removed and returned to the microtome, then it must be re-aligned to the knife
as instructed.
35.
Or
return to the block to cut more thick sections and stain them as described.
36.
When
finished with all thick sectioning, the diamond Histo knife must be cleaned
before storing away. To do this:
a)
First,
remove any remaining floating sections in the boat water by wiping the water
surface with a Ross lens tissue.
b)
Remove
water from the boat with a plastic pipette or syringe.
c)
Use
a clean, razor-beveled end of a Styrofoam stick (usually provided by the
manufacturer) that has been lightly soaked in 100% ethanol to do the cleaning
as described in the next step.
d)
Gently
and lightly, sweep the beveled Styrofoam end across the length of the diamond
knife, starting about 1 mm from the edge, from left to right. Then repeat from
right to left. See diagram:

e)
Bring
the knife over to the 2X-distilled faucet and rinse the knife under 2X
distilled running water.
f)
Examine
the knife-edge under the stereomicroscope of the microtome for any residual
debris.
g)
If
any, then re-clean as directed. See these same instructions from the diamond
knife manufacturer’s manual, Diatome, US, e.g.
h)
If
completely finished, then store the diamond knife in its box.
i)
Remember
to turn off the compressed N2 to the air table!
SPECIAL
PROCEDURES
Tissue
measurement in Epon flat blocks
1)
Place
the untrimmed or trimmed flat block, tissue side up, on a glass slide under the
light microscope (Nikon Labophot-2).
2)
Turn
on the microscope’s light source and focus on the tissue under 10X objective.
Position the block so that the length of tissue can be measured with the
eyepiece reticule (focus this eyepiece reticule).
3)
To
enhance illumination of the tissue, shine the fiber optic light source near it.
One can see right through the hippocampal CA1 pyramidal cell body layer in a
100-micron thick or less tissue slice. This reference is advantageous in
measuring distance from the cell body layer. See diagram:

4)
The
magnification values under the objective lenses are as follows:
Plan 2: each small division
is equal to 50 microns
Plan 10: each small division
is equal to 10 microns
Plan 20: each small division
is equal to 5 microns
Plan 40: each small division
is equal to 2.5 microns
5)
The
Epon block can continued to be measured in this way after thick and thin
sectioning. In addition, the block can be photographed. See protocol on “Light Microscopy
Photography”.